Abstract

This Research Paper constitutes the experimental investigation of corrosion test by using artificial Saliva and Hydroxiapatite plasma spray Coating on SS316L of the existing material used for orthopedic implant i.e. for Femur prosthesis. in this the specimen SS316L is dipped or immersed in artificial Saliva for one week at room temperature in accordance with ASTM G-31and analysis is carried out that there is no weight loss in the specimen and it indicates that there is no corrosion in the specimen and also the Hydroxyapatite plasma spray coating is done on this specimen from the experimental results the coating is withstand it is found that thickness of coating is around 50-70µm and hence we suggest this Bio-material to use for femur bone prosthesis.